GDS-2000A Series User Manual 
144
 
Trigger 
The trigger configures the conditions for when the GDS-2000A 
captures a waveform.  
The following trigger overview only applies to the analog channels, 
for triggering details using the optional logic analyzer module, 
please see the GDS-2000A Options User Manual for details.  
Trigger Type Overview 
 
Edge 
The edge trigger is the simplest trigger type. An 
edge trigger triggers when the signal crosses an 
amplitude threshold with either a positive or 
negative slope.  
 
 
 
Rising edge trigger 
 
 
 
Falling edge trigger 
Delay 
The Delay trigger works in tandem with the edge 
trigger, by waiting for a specified time or number 
of events before the edge trigger starts. This 
method allows pinpointing a location in a long 
series of trigger events. 
Note: when using the delay trigger, the edge trigger 
source can be any one of the channel inputs, the EXT 
input or the AC line.  
 
Delay trigger example (by event) 
 
1
2
3
A
B
C
D
 
A  Ext. trigger input 
(Delay trigger) 
 
B  Source (Edge trigger) 
 
C  Delay event count (3) 
 
D  First triggering point