Lecroy WAVEACE 2024 4-channel oscilloscope, Digital Storage oscilloscope, WaveAce 2024 Manual Do Utilizador
Códigos do produto
WaveAce 2024
Operator's Manual
920836 Rev A
31
ACQUISITION - WAIT
Description
The WAIT command prevents your instrument from analyzing new commands until the current
acquisition has been completed. The optional argument specifies the timeout (in seconds) after
which the scope stops waiting for new acquisitions. If <t> is not given, or if <t> = 0.0, the scope
waits indefinitely.
acquisition has been completed. The optional argument specifies the timeout (in seconds) after
which the scope stops waiting for new acquisitions. If <t> is not given, or if <t> = 0.0, the scope
waits indefinitely.
Command Syntax
WAIT [<t>]
<t>:=
timeout in seconds (default is indefinite)
Example
ARM;WAIT;*OPC;C1:PAVA? MAX
This example finds the maximum amplitudes of several signals acquired one after another. ARM
starts a new data acquisition. The WAIT command ensures that the maximum is evaluated for
the newly acquired waveform.
starts a new data acquisition. The WAIT command ensures that the maximum is evaluated for
the newly acquired waveform.
C1:PAVA? MAX instructs the oscilloscope to evaluate the maximum data value in the Channel 1
waveform.
waveform.
Related Commands
*TRG, TRIG_MODE, ARM