Lecroy WaveStation 2022 Function Generator, Frequency Generator WaveStation 2022 Manual Do Utilizador
Códigos do produto
WaveStation 2022
WaveStation SCPI Command Reference Manual
920595 RevA
14
CARR, MEAN
<mean>
Only applicable when the carrier waveform is set to Noise, this parameter sets
the mean.
the mean.
Note: Tables specifically covering command arguments are listed in the order in which they are functionally
processed for convenience.
processed for convenience.
Where:
<state>:= {ON,OFF}
<period>:= {The default unit is in “S”. 1µs to 500s}
<start phase>:= {0 to 360}
<gate ncycle>:= {GATE,NCYC}
<trigger>:= {EXT,INT,MAN}
<delay>:= {The default unit is in “S”. 0s to 500s}
<polarity>:= {NEG,POS}
<trig mode>:= {RISE,FALL,OFF}
<edge>:= {RISE,FALL}
<cycle time> := {1 to 50000 cycles}
<wave type>:= {SINE,SQUARE,RAMP,PULSE,NOISE,ARB}
<frequency>:= {The default unit is in “HZ”. The minimum value is 1xe-6 HZ and
the maximum depends on the WaveStation model.}
the maximum depends on the WaveStation model.}
<amplifier>:= {The default unit is in “V”. The C1 minimum value is 0.004V,
maximum is 6V. The C2 minimum value is 0.004V and the maximum is 20V.}
maximum is 6V. The C2 minimum value is 0.004V and the maximum is 20V.}
<offset>:= {The default unit is in “V”. The maximum value depends onyou’re
your WaveStation model.}
your WaveStation model.}
<duty>:= {If the waveform type is set to Square, the range is from 20% to 80%. If
the waveform type is set to Pulse, the range is from 0.1% to 99.9%}
the waveform type is set to Pulse, the range is from 0.1% to 99.9%}
<symmetry>:= {0% to 100%}
<carr delay>:= {The maximum delay is the Pulse period, and the minimum value
is 0. Units are in S.}
is 0. Units are in S.}
QUERY
SYNTAX
<channel>: BursTWaVe?
RESPONSE
FORMAT
<channel>: BursTWaVe <type>,<state>,<period>
EXAMPLE
1
Set the C1 Burst waveform period to 1S.
C1: BTWV PRD, 1S
EXAMPLE
2
Set the C1 Burst waveform delay to 0S.
C1: BTWV DLAY, 0S