Lecroy WS64MXs-B 4-channel oscilloscope, Digital Storage oscilloscope, WS64MXs-B Ficha De Dados

Códigos do produto
WS64MXs-B
Página de 6
 
 
LeCroy Corporation 
WaveAce Integration with Wavestation 
page 
|
 
3 of 6 
 
Transfers from WaveAce to WaveStation 
Getting signals from WaveStation into WaveAce is a simple matter of connecting the output of the signal 
generator to the input of the oscilloscope via coaxial cable. Transferring signals from WaveAce to WaveStation is 
even simpler once the USB connections with the PC are established. The following is a step-by-step procedure 
for sending a real-world signal from the WaveAce oscilloscope to the WaveStation arbitrary signal generator. The 
procedure assumes correct installation of all necessary software and hardware drivers. 
 
1.  Connect both instruments to the PC via USB cables. Turn on both of the instruments. 
 
2.  For this demonstration, a WaveAce 2034 oscilloscope and LeCroy’s WaveSource 100 oscilloscope 
evaluation signal source were used to generate and acquire a pulsed signal as shown in Figure 3. Of 
course, any signal of interest may be acquired. A WaveStation 2052 served as the waveform generator. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Figure 3:  A screen capture of the WaveAce oscilloscope shows the acquired signal from 
 
 
the WaveSource 100 evaluation source 
 
3.  On the PC, open National Instruments’ Measurement & Automation Explorer application. The application 
will automatically detect the WaveAce as shown in Figure 4. Note that the oscilloscope’s USB port must 
be set to USBTMC. This is accomplished in the oscilloscope’s Utility menu.