Lecroy WSXS-SPIBUS TD WSXS-SPIBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® oscilloscopes. WSXS-SPIBUS TD Manual Do Utilizador

Códigos do produto
WSXS-SPIBUS TD
Página de 204
Operator's Manual 
919586 RevA 
189 
Additional Dialog Controls Based on Packet Category Selections 
 
With the Any Sync category selected, no additional controls are provided on the dialog. 
 
With the Token category selected, up to three Token PID controls may be used together (using OR 
operator logic) unless Any is selected on the first PID, disabling the other two. Token PID controls (when 
enabled) can be set to AnyInOutSetupPingSplitSOF , or Preamble
 
PLEASE NOTE THE FOLLOWING:  
 
Token PID values are enabled or disabled based on Bus Speed selection. 
 
When Any and Preamble Token PID values are chosen, no additional controls are provided for that 
specific Token PID. 
 
When InOutSetup, and Ping Token PID values are chosen, additional controls for DevAddress 
and End Point are provided for that specific Token PID. 
 
When the Split Token PID value is chosen, additional controls for Hub AddressSSCEPort, and 
ET are provided for that specific Token PID. 
 
When the SOF Token PID value is chosen, a Frame Number control is provided for that specific 
Token PID. 
 
With the Data category selected, up to three Data Pattern PID controls may be used together (using OR 
operator logic) unless Any Type is selected on the first PID, disabling the other two. Data Pattern PID 
controls (when enabled) can be set to Any TypeData0Data1Data2, or MData
 
PLEASE NOTE THE FOLLOWING:  
 
Data Pattern PID values are enabled or disabled based on Bus Speed selection. 
 
Compare Type (select Data Value or Data Length), Compare Operator (select equal or unequal), At 
position
, and Data (1-16 Bytes) (in Hexadecimal or Binary based on selection from step 2 
previously shown) controls are provided for each Data Pattern PID. 
 
With the Handshake category selected, Handshake Type control buttons are available for Any TypeACK
NAKSTALLNYET, and Error on the dialog. 
 
With the User Define category selected, a Data (1-16 Bytes) control is provided for entering a specific 
value in Hexadecimal or Binary (based on selection from step 2 previously shown).