Lecroy WSXS-SPIBUS TD WSXS-SPIBUS TD trigger and decoder extension Compatible with LeCroy WaveSurfer® oscilloscopes. WSXS-SPIBUS TD Manual Do Utilizador

Códigos do produto
WSXS-SPIBUS TD
Página de 204
Operator's Manual 
919586 RevA 
55 
 
Viewing - Select to view the protocol data in BinaryHexadecimal, or ASCII modes. 
Note: If the trigger is set up first, the trigger settings copy into the decode settings.  
 
Include R/W Bit - Some engineers think of the 7-bit address pattern as including the R/W bit (i.e. 8-bits) 
and others think of the address pattern as not including the R/W bit (i.e. 7-bits). If you decoded I
2
messages include 7-bit addresses, mark the checkbox if you want to include the R/W bit in the decoded 
Address value. 
Note: There is an identical checkbox selection in the I
2
C trigger setup dialog. These two setups are 
dynamically linked, so selections here in decode results in an identical selection in trigger. This ensures 
that the decode address format matches trigger setup information. 
I
2
C Levels Right-Hand Dialog 
The Levels right-hand dialog provides detailed controls and setup conditions as follows: 
 
 
Level Type and Vertical Level - The message decoding algorithm setup is performed here. The level is 
normally set up in %, and defaults to 50%. To adjust the level, touch inside the number area to highlight 
the box title, and then use the oscilloscope front panel Adjust knob to adjust. Or touch inside the number 
area twice and select a value using the pop-up numeric keypad. 
PLEASE NOTE THE FOLLOWING:  
 
The set Level appears as a dotted horizontal line across the oscilloscope grid. 
 
If initial decoding indicates that there are a number of error frames, make sure that the level is set 
to a reasonable value. 
 
DATA and CLOCK can have different level settings, but they are typically the same level. 
Creating an I
2
Cbus Trigger Condition 
The following trigger setup detail topics show the dialog selections for an I
2
Cbus Trigger with detail on some of 
the setup conditions. 
Note: Refer to Using The D Supported Protocol Toolsets (on page 14) to correctly access the Trigger Condition 
dialog specific to your desired protocol.