Select condition values by touching fields (using your finger, or use a mouse pointer). A pop-up is shown where 
you can choose from EqualNot EqualLess thanLess than or Equal toGreater thanGreater than or Equal to
In Range, or Out Range condition values. 
LINbus Trigger Setup Detail 
The following topic provides specific control settings for a LINbus Trigger. 
 
The previously numbered LINbus trigger sections correspond with the following explanations. 
1.
 
S
OURCES 
S
ETUP
 
 
DATA - The pop-up dialog is used to select the appropriate channel or EXT input for each. Set this field up 
with caution or your trigger may not function correctly. 
 
Threshold (Trigger) - Adjust the vertical level for the trigger. Much like an Edge trigger, a user must 
specify the level used in order to process the incoming signals and determine whether the desired serial 
data pattern is meeting the set trigger condition. 
2.
 
LIN
 
S
ETUP
 
 
Bitrate - The LIN trigger can be configured to trigger on LIN busses at several different bitrates including 
1.2 kb/s2.4 kb/s4.8 kb/s9.6 kb/s10.417 kb/s, and 19.2 kb/s
3.
 
T
RIGGER 
T
YPE
 
The LIN trigger can be configured to trigger on simple Start of Frame (Break) conditions, ID only, or 
complete ID+DATA conditions with DATA conditions other than equals. Some Error Frame triggering is 
also supported. 
Choose a desired Trigger Type and the trigger dialog changes based on the selection made. 
For example, the following trigger selections disable or enable fields on the Setup FormatFrame ID 
Setup
Data Pattern Setup, and Checksum Error Setup... (Checksum error only for Error trigger type) 
sections of the trigger dialog: 
 
Break - When selected, the Setup Format, Frame ID Setup, and Data Pattern Setup fields are disabled. 
 
Frame ID - When selected, Setup Format and Frame ID Setup fields are enabled, and the Data Pattern 
Setup fields are disabled. 
 
ID + Data - When selected, Setup Format, Frame ID Setup, and the Data Pattern Setup fields are enabled. 
 
Error - When selected, Setup Format fields are enabled and the Frame ID Setup fields are disabled. The 
Pattern Setup fields (step 6a, as follows) aren't shown (and are therefore disabled). Instead, the 