EXTERNAL CONTROL
101
Using the SIGNAL I/O to Start/Stop Tests
Background
To use the SIGNAL I/O port the Start Ctrl
settings have to be set to SIGNAL I/O in the
Common Utility menu.
settings have to be set to SIGNAL I/O in the
Common Utility menu.
Panel operation
1. Set the Start Ctrl option to SIGNAL
I/O.
Page 76
2. Turn off the power.
3. Connect the Input/Output signals
to the SIGNAL I/O port.
4. Turn the tester back on.
5. To start the testing, short the
INPUT_STOP and INPUT_COM
line for a minimum of 1ms to put
the tester into READY status.
line for a minimum of 1ms to put
the tester into READY status.
6. To start the testing, short the
INPUT_START and INPUT_COM
lines for a minimum of 1ms.
lines for a minimum of 1ms.
7. To stop the testing, temporarily
short the INPUT_STOP and
INPUT_COM line again.
INPUT_COM line again.