Insulation Testers 
 
Measuring Polarization Index and Dielectric Absorption Ratios (Model 1507)   
  
15 
• 
The primary display shows ---- until you press the 
T
 button and a valid resistance reading is 
obtained. 
• 
The high voltage symbol (
Z
) along with a primary 
display of >30 V warns if voltage greater that 30 
V ac or dc is present. If high voltage is present, 
the test is inhibited. 
5.
 
Press and release 
T
 to start the test. During 
testing, the secondary display shows the test voltage 
applied to the circuit under test. The high voltage 
symbol (
Z
) along with a primary display showing the 
resistance in M
Ω
 or G
Ω
. The 
t
 icon appears on the 
lower portion of the display until the test is finished.  
When the test is completed, the PI or DAR value is 
displayed on the primary display. The circuit under 
test will automatically be discharged through the 
Tester. If either value used to calculate PI or DAR 
was greater than the maximum display range, or the 
1-minute value was greater than 5000 M
Ω
, the 
primary display will show 
Err
• 
When resistance is higher than the maximum 
display range, the Tester displays the > symbol 
and the maximum resistance for the range. 
• 
To interrupt a PI or DAR test before it is 
completed, momentarily press 
T
. When you 
release 
T
, the circuit under test will 
automatically be discharged through the Tester.