Greisinger Oxy 3610 602144 Manual Do Utilizador

Códigos do produto
602144
Página de 6
T02.0.3X.6C-05 
 
Page 1 of 6  
Operating Manual  
for Dissolved Oxygen-Measuring transducer  
OXY 3610 MP 
Specification: 
Measuring range: 
0.00 to 25.00 mg/l dissolved oxygen 
Output signal: 
refer to type plate   
Accuracy:  (at nominal temperature) 
  Display 
±1.5% of measured value ±0.2 mg/l 
  Output signal: 
±0.2 % FS 
Connection: 
4 - 20 mA  (2-wire) 
Voltage  (3- resp. 4-wire) 
Electric isolation: 
input electrically isolated 
Auxiliary energy: 
(supply voltage)
  Uv = 12 - 30 V DC (4-20mA) 
Uv = 18 - 30 V DC (0-10V)   
or refer to type plate 
Reverse voltage protection:  50V permanent 
Perm. impedance 
(at 4-20mA)
:    R
A
(Ohm) < ( (Uv - 12V) / 0.02A ) 
Example: for Uv = 18V: 
R
A
 < (18V - 12V) / 0.02A     =>   R
A
 < 300 Ohm
 
Permissible load 
(at 0-...V)
:   
R
L
(Ohm) > 3000 Ohm 
Electrode: 
Special oxygen electrode GWO3600MU (in scope of supply!) 
Electrode type:  
active diaphragm type, with integrated NTC resistor 
Response time: 
95% in 10 sec., depending on temperature 
Operation life: 
3 years er more, depending on proper maintenance 
Operating pressure: 
max. 3 bar. 
Mounting diameter:   
Ø 12,0 ±0,2 mm (also suitable for ½" glanding)  
Length:   
approx. 220 mm (incl. anti-buckling glanding), Mounting length: approx. 110 mm 
Weight  
approx. 180 g 
Electrode connection:   
5 pole screwable socket  
Temperature compensation:   automatic, 0 ... 50 °C  
Calibration:  
via keypress at ambient air 
Display:   
approx. 10 mm high, 4-digit LCD-display 
Nominal temperature: 
20 °C 
Operating temperature: 
0 to 50 °C (electrode 0 to 40 °C) 
Relative humidity: 
0 to 95 %RH (non-condensing) 
Storage temperature: 
-20 to 70 °C (electrode 0 to 60 °C) 
Housing: 
ABS (IP65 - with the exception of electrode and temperature probe connection sockets) 
Dimensions: 
82 x 80 x 55 mm (without elbow-type plug and sensor sockets) 
Mounting: 
With fixing holes for wall mounting (in housing - accessible after cover has been 
removed). 
Mounting distance: 
50 x 70 mm, max. shaft diameter of mounting screws is 4 mm. 
Electric connection: 
elbow-type plug conforming to DIN 43650 (IP65), 
max. wire cross section: 1.5 mm², wire/cable diameter from 4.5 to 7 mm 
EMC: 
The device corresponds to the essential protection ratings established in the Regulations of the 
Council for the Approximation of Legislation for the member countries regarding electromagnetic 
compatibility (2004/108/EG). Additional error: <1% 
WEEE-Reg.-Nr. DE93889386 
GREISINGER 
electronic GmbH
 
D - 93128 Regenstauf, Hans-Sachs-Straße 26 
 +49 (0) 9402 / 9383-0   +49 (0) 9402 / 9383-33   