PCS-1000 Specifications 
The specifications apply when the PCS is powered on for at least 30 
minutes.  
General  
 
Power Supply 
100 V / 120 V / 220 V / 240 V ±10% 
Power Line Frequency  50/60Hz 
Operating Environment  Full accuracy for 0˚C to 50˚C, Full accuracy to 80% R.H. 
at 40˚C 
Storage Environment 
-40˚C to 70˚C 
Power Consumption 
Max 35VA 
Dimensions 
210mm (W) * 80mm (H) * 390mm (D) 
Weight 
Approximately 6 kg 
DC Characteristics  
 
DC Voltage 
Range 
1 Year 
23˚C ± 5˚C  
Temperature 
Coefficient/˚C 
200.0000 mV 
0.0050 + 0.0035 
0.0005 + 0.0005 
2.000000 V 
0.0050 + 0.0010 
0.0005 + 0.0001 
20.00000 V 
0.0050 + 0.0010 
0.0005 + 0.0001 
200.0000 V 
0.0050 + 0.0010 
0.0005 + 0.0001 
1000.000 V 
0.0050 + 0.0020 
0.0005 + 0.0001 
Accuracy specification : ± (% of reading + % of range) 
Voltage input Resistance: 10MΩ for all DC Voltage ranges 
 
 
DC Current 
Range 
Burden  
Voltage 
1 Year  
23˚C ± 5˚C  
Temperature 
Coefficient/˚C 
 
30.00000 mA 
<0.4 V 
0.01 + 0.005 
0.001 + 0.002 
 
300.0000 mA 
<0.5 V 
0.01 + 0.005 
0.001 + 0.002 
 
3.000000 A 
<0.8 V 
0.01 + 0.005 
0.001 + 0.002 
 
30.00000 A 
<0.8 V 
0.01 + 0.005 
0.001 + 0.002 
 
300.0000 A 
<0.8 V 
0.02 + 0.005 
0.001 + 0.002 
 
Accuracy specification : ± (% of reading + % of range)