Calibration 
In order to compensate for ageing of the sensor and air pressure fluctuations due 
to weather changes, the sensor has to be calibrated at regular intervals. The 
sensor will be calibrated to the atmospheric oxygen concentration of 20.95%. 
We recommend to calibrate directly before starting the measuring process.  
Calibration:
.
The sensor has to be subjected to air (make sure that rooms are 
thoroughly aired). 
Press the 
 key for 2 sec till 
 appears. 
The calibration will be automatically completed as soon as the measuring value is 
stable (takes a few seconds).  Afterwards the rating of the sensor state will be 
shown for a short time. 
If an error message (CE.3, CE.4, CE.6) is displayed, the sensor signal is 
invalid. Press any key to restart, the previous calibration will be restored. 
The display is blinking afterwards until a valid calibration is performed. 
10  Sensor rating 
The rating of the sensor state is calculated and saved after successful calibration.  
Watch sensor rating: Press shortly the 
 
key, the display shows for a short 
time i.e. 
 = 100%. 
The rating is displayed in 10 percent steps: 100% = optimal sensor status. Lower values are 
indicating that the sensor life time will soon be reached. (50% doesn‟t mean 50% of the life 
time, but 50% of the reference signal!). Erroneous P.AbS entries may cause to low ratings
11  Errors and system messages 
no  
display 
The battery is used up and has to be exchanged, or the device is defective 
blinking 
display 
The last calibration was not successful. A new calibration has to be performed! 
 
The battery is used up, for a short period further measurements are possible 
 
The battery is used up and has to be replaced, measurements are no more 
possible  
 
Calibration error: sensor voltage is too low (sensor used up or wrong chosen air 
pressure) 
 
Calibration error: sensor voltage is too high (wrong chosen air pressure or 
sensor defective (membrane crack, etc.) 
 
Calibration error: signal not stable 
 
Measuring error: range has been exceeded 
 
Measuring error: measuring value has fallen below permitted range 
 
System error: the device is defective or considerably out of the allowed ambient 
temperature range 
 
Value could not be calculated (e.g. input of MOD-calculation is error)