Keysight Technologies U1168A Ficha De Dados

Página de 8
5
Electrical Specifications
Measuring rate
Function
Times/second
ACV
7
DCV (V or mV)
7
Ω
14
Diode
14
Capacitance
4 (< 100 µF)
DCA (μA, mA, A)
7
ACA (μA, mA, A)
7
Temperature
7 (single)
Frequency
1 (> 10 Hz)
General Specifications
Display
Dual display (secondary display is intended for temperature function display only) 
consists of 4-digit liquid crystal display (LCD) with maximum reading of 11,000 counts. 
Automatic polarity indication.
Power consumption
0.22 VA maximum
Battery type and life
Four single standard 1.5 V AAA batteries (Alkaline or Zinc Chloride type); 300 hours 
typical
Operating environment
•  Full accuracy at –10 to 55 °C; and to 80% RH for temperatures up to 30 °C, decreasing 
linearly to 50% RH at 55 °C
•  0 – 2000 meters per IEC 61010-1 2nd Edition CAT III, 1000 V / CAT IV, 600 V IEC 
61010-1 2
nd
 Edition
Storage compliance
–20 to 70 °C
Safety compliance
•  IEC 61010-1:2001 / EN61010-1:2001
•  Canada: CSA C22.2 No. 61010-1:2004
Measurement category
CAT III 1000 V / CAT IV 600 V Overvoltage Protection, Pollution Degree 2
EMC compliance
• CISPR 
11:1990/EN55011:1990
• Canada: 
ICES-001:2004
•  Australia/New Zealand: AS/NZS CISPR11:2004
Common Mode Rejection Ratio (CMRR)
> 90 dB at DC, 50/60 Hz ± 0.1% (1 kΩ unbalanced)
Normal Mode Rejection Ration (NMRR)
> 60 dB at 50/60 Hz ± 0.1%
Crest factor
< 3.0
Temperature coefficient
0.1 × (specified accuracy) / °C (from –10 to 18 °C or 28 to 55 °C)
Shock and vibration
Tested to IEC/EN 60068-2
Dimensions (H x W x D)
193.8 mm x 92.2 mm x 58.0 mm
Weight
•  450 g with batteries
•  400 g without batteries
Warranty
•  Three years for main unit
•  Three months for standard shipped accessories