Keysight Technologies U2652ADigital USB input module U2652A OPT 902 + ABD Ficha De Dados

Códigos do produto
U2652A OPT 902 + ABD
Página de 75
26
Product specifi cations
Model number
U2531A
U2541A
U2542A
Analog input
Resolution
14 bits
16 bits
Number of channels
4 differential input channels (software selectable/channel)
Maximum sampling rate
2 MSa/s
250 kSa/s
500 kSa/s
Programmable bipolar input range
[1]
±10 V, ±5 V, ±2.5 V, ±1.25 V
Programmable unipolar input range
0 to 10 V, 0 to 5 V, 0 to 2.5 V, 0 to 1.25 V
Input coupling
DC
Input impedance
1 GΩ/100 pF
Operational common mode voltage range
±8.0 V maximum
Overvoltage range
Power-on: Continuous ±30 V, Power-off: Continuous ±15 V
Trigger sources
External analog/digital trigger, SSI/star trigger
[2]
Trigger modes
Pre-trigger, delay-trigger, post-trigger, and middle-trigger
FIFO buffer size
Up to 8 MSa
Analog output
Resolution
12 Bits
Number of channels
2
Maximum update rate
1 MSa/s
Output ranges
0 to 10 V, ±10 V, 0 to AO_EXT_REF, ±AO_EXT_REF
[3]
Output coupling
DC
Output impedance
0.1 Ω Typical
Stability
Any passive load up to 1500 pF
Power-on state
0 V steady state
Trigger sources
External analog/digital trigger, SSI/star trigger
[2]
Trigger modes
Delay trigger, post trigger
FIFO buffer size
1 Channel used: Maximum 8 MSa
4 Channels used: Maximum 2 MSa/ch
Glitch energy
5 ns-V (Typical), 80 ns-V (Maximum)
Driving capability
5 mA
Function generation mode
Sine, square, triangle, sawtooth, and noise waveforms
Digital input/output
Number of bits
24-bit programmable input/output
Compatibility
TTL
Input voltage
VIL = 0.7 V maximum; IIL = 10 μA maximum
 VIH = 2.0 V minimum; IIH = 10 μA maximum
Input voltage range
–0.5 V to +5.5 V
Output voltage
VOL = 0.45 V maximum; IOL = 8 mA maximum
 VOH = 2.4 V minimum; IOH = 400 μA maximum
General purpose digital timer/counter
Maximum count
(231 – 1) bits
Number of channels
2 Independent up/down counter
Compatibility
TTL
Clock source
Internal or external
Base clock available
48 MHz
Maximum clock source frequency
12 MHz
Input frequency range
[4]
0.1 Hz to 6 MHz at 50% duty cycle
Pulse width measurement range
0.167 μs to 178.956 s ±0.0833 μs