DC characteristics
Accuracy specifi cations
[1]
Model
U2541A, U2542A
Analog input
Unipolar range (V)
Offset error (mV)
[2]
Gain error (mV)
Accuracy (% of reading + offset error)
[3]
10
1.0
1.0
0.02% + 1.0 mV
5
1.0
1.0
0.04% + 1.0 mV
2.5
1.0
1.0
0.08% + 1.0 mV
1.25
1.0
1.0
0.16% + 1.0 mV
Bipolar range (V)
10
1.0
2.0
0.02% + 1.0 mV
5
1.0
1.0
0.02% + 1.0 mV
2.5
1.0
1.0
0.04% + 1.0 mV
1.25
1.0
1.0
0.08% + 1.0 mV
Model
U2531A
Unipolar range (V)
Offset error (mV)
[2]
Gain error (mV)
Accuracy (% of reading + offset error)
[3]
10
2.0
3.0
0.06% + 2.0 mV
5
1.5
1.5
0.06% + 1.5 mV
2.5
1.0
1.0
0.08% + 1.0 mV
1.25
1.0
1.0
0.16% + 1.0 mV
Bipolar range (V)
10
2.0
6.0
0.06% + 2.0 mV
5
1.5
3.0
0.06% + 1.5 mV
2.5
1.0
2.0
0.08% + 1.0 mV
1.25
1.0
1.0
0.08% + 1.0 mV
Model U2541A, 
U2542A
Analog output
Unipolar range (V)
Offset error (mV)
[2]
Gain error (mV)
Accuracy (% of reading + offset error)
[4]
10
1.0
2.0
0.02% + 1.0 mV
Bipolar range (V)
10
1.0
2.0
0.02% + 1.0 mV
Model
U2531A
Unipolar range (V)
Offset error (mV)
[2]
Gain error (mV)
Accuracy (% of reading + offset error)
[4]
10
1.0
3.0
0.03% + 1.0 mV
Bipolar range (V)
10
1.0
3.0
0.03% + 1.0 mV
[1]
 Specifi cations are based on 20 minutes warm-up, and self-calibration temperature at 23 °C.
[2]
 Offset error is measured at 0 V.
[3]
 Accuracy = ± [% of Gain error/(Measured value – Midscale) + Offset error]
[4]
 Accuracy = ± (% of Gain error/Output value + Offset error)