Keysight Technologies U2652ADigital USB input module U2652A OPT 902 + ABD Ficha De Dados

Códigos do produto
U2652A OPT 902 + ABD
Página de 75
61
[1]
 Frequency measurement can only be done in auto range mode. Specifi cations are for 30 minutes warm-up time, using one second aperture. Measuring 
method is using reciprocal counting technique with AC coupled input at AC voltage function. Gate time of 0.1 second or 1 second.
[2]
 Only applicable for square wave measurement.
Frequency specifi cations
[1]
Frequency accuracy
Function
Range
Accuracy
(% of reading + % of 
range)
Minimum input 
frequency
Temperature 
coeffi cient (% of range)
Frequency
20 Hz to 300 kHz
0.0200 + 0.003
1 Hz
0.005
Frequency Sensitivity for AC Voltage
Function
Range
Minimum sensitivity (RMS sine wave)
Frequency (Hz)
20 ~ 100k
100k ~ 300k
AC voltage
100 mV
[2]
20 mV
20 mV
1  V
100 mV
120 mV
10  V
1 V
1.2 V
100  V
10 V
20 V
250 V
100 V
120 V
Temperature specifi cations
Temperature accuracy
Function
Thermistor type
Range
Accuracy
Temperature coeffi cient
Temperature
5 kΩ thermistor
–80.0 °C to 150 °C
Probe accuracy + 0.2 °C
0.002 °C
–112 °F to 302 °F 
NOTE
To ensure better measurement results and to guard against the change of environment or setup, always 
enable the Null offset.
Typical Reading Speed (in seconds) Characteristics
Test/Range
100 mV(20 Hz)
1V(20 Hz)
10V(20Hz)
100V(45 Hz)
300V(45 Hz)
ACV
0.979
0.979
0.978
0.979
0.979
Test/Range
10 mA
100 mA
1 A
2 A
ACI
0.979
0.979
0.979
0.979
Freq
1.190