Keysight Technologies U1560A Ficha De Dados

Página de 11
7
Measurement Characteristics
Full scale reading
6,000-count
DC voltage, True RMS AC voltage
Maximum input voltage, 600 Vrms CAT II, 300 Vrms CAT III
DC coupled input coupling
Continuity
Beeper < 60 W in 600 W range
Data Logger
Source
Digital multimeter measurements
Range
10 divisions
Record size
Up to 8800 data points (with option 001)
Time span
Auto range 150 seconds to 20 days
Time reference
Time from start
Record method
Selectable minimum, maximum and average
Display System
Display
4.5-inch diagonal color CSTN LCD
Resolution
320 x 240 pixels
Control
Contrast control, infi nite persistance on/off
Built-in help system
Functional help displayed by pressing help button
Real-time clock
Time and date (user-adjustable)
Storage
Save/Recall (non-volatile)
Up to 10 setups and traces
Auxiliary Meter Specifi cations ± (% of reading + % of range)
Function
Range
Frequency
1 year Tcal ± 5 ºC
Temperature
[2]
,
 
ºC
600.0 ºC 
0.3 + 0.08
6000 ºC
0.3 + 0.08
Temperature
[2]
, ºF
600.0 ºF
0.3 + 0.08
6000 ºF
0.3 + 0.08
AC Current
[3]
60.00 A
50 Hz – 1 kHz
1.0 + 0.2
600.0 A
50 Hz – 1 kHz
1.0 + 0.2
[2] U1586B temperature module is needed to make measurements and is available as option.
[3] U1583B AC current clamp needed to make measurements and is available as option.