Bk Precision BK-8500 Electronic Load 1 mA - 30 A 0.1 - 120 Vdc 0 - 300 W BK-8500 Ficha De Dados

Códigos do produto
BK-8500
Página de 76
Lvl
Menu item
Function
2
BAUDRATE SET
Set the serial bus connection speed.
3
4800
3
9600
3
19200
3
38400
2
COMM. PARITY SET
Serial connection parity.
3
NONE
3
EVEN
3
ODD
2
ADDRESS SET
Set the instrument's address.  Must be an integer 
between 0x00 and 0xFE inclusive.
2
KEY LOCK SET
Set the instrument's password to 1 - 4 digits.  Press the 
Enter key without any data entry to remove the 
password. If you accidentally locked the keyboard, 
follow instructions at the end of this manual, chapter “in 
case of trouble” to unlock the keypads.
2
EXIT
Return to prior menu level.
1
SYSTEM SET
2
MAX CURRENT SET
Set the maximum allowed current.  Exceeding this 
value will cause the load to shut off.  This value also 
becomes the maximum current you can set.
2
MAX POWER SET
Set the maximum allowed power.  Exceeding this value 
will cause the load to shut off.  This value also becomes 
the maximum power you can set.
2
MAX VOLTAGE SET
Set the maximum allowed voltage.  Exceeding this 
value will cause the load to shut off.  This value also 
becomes the maximum voltage you can set.
2
VOLTAGE ON SET
Used to set an ON voltage threshold.  See the Voltage 
threshold
 section.
2
VOLTAGE OFF SET
Used to set an OFF voltage threshold.  See the 
Voltage threshold section.
2
EXIT
Return to prior menu level.
1
LIST SET
Lists are programmed sequences of load values.
2
MODE SET
3
FIXED MODE
Load characteristics set by front panel.
3
LIST MODE
Load characteristics are dynamic and controlled by a 
list stored in memory.
2
CALL LIST FILE
Recall an existing list from nonvolatile memory.
3
RECALL N
N is list number.
2
EDIT LIST FILE
Note that each list file has four lists, one for each 
operation mode (CC, CV, CW, CR).
3
CURRENT LIST
4
ONCE
List is executed once per trigger.
8500 DC Load Series                    
Version:  September 2, 2009 
Page 13 of 76