Gw Instek PEL-3211 Electronic Load PEL-3211 Ficha De Dados
Códigos do produto
PEL-3211
PEL-3000 User Manual
90
5. Set the Delay Time.
The delay time setting will delay activating the
Go-NoGo testing by a specified amount of time.
Go-NoGo testing by a specified amount of time.
The delay setting can compensate for startup
oscillation and other instabilities during startup.
oscillation and other instabilities during startup.
Delay Time
0.0~1.0 seconds (0.1s resolution )
Note
When the Main settings are saved or recalled, the
Go-NoGo settings are also saved/recalled. See the
Save/Recall chapter for details, page 119.
Go-NoGo settings are also saved/recalled. See the
Save/Recall chapter for details, page 119.
Running a Go-NoGo Test
Description
Go-NoGo test results are displayed in the
measurement panel.
measurement panel.
GO indicates pass (good).
NG indicates fail (no good).
Operation
1. Press
Main
> Configure[F5] > Go-NoGo[F3].
2. Set SPEC Test to ON.
When SPEC Test is ON, SPEC will appear in the
Operation Status Panel. This means the unit is
ready for Go-NoGo testing.
Operation Status Panel. This means the unit is
ready for Go-NoGo testing.
3. Turn the load on.
The test starts from the time the load was turned
on + the Delay Time.
on + the Delay Time.