Figure 12
Figure 12 shows a possible progression of a set value 
(U, I, P or R) with adjustable pulse widths and variable 
amplitude. The  rise/fall  time  is  also  adjustable,  but  is 
equal for A and B. 
If the rise/fall time is set to minimum, the signal of the 
pulsed operation is a nearly ideal square wave. Figure 
12  is  only  a  clarifying  view. The  real  progression  of, 
for  example,  an  input  current  which  is  pulsed  with  a 
frequency of 1kHz, will differ more or less. It depends 
on many circumstances, like the regulation time of the 
feeding power supply, the regulation time of the load, line 
impedances etc.
In „pulsed“ operation the actual values are also pulsing 
and the values constantly change in the display, so they 
are  not clearly readable.Thus the actual values are now 
only indicated at pulse times >=1s (A and B each) and 
can be read quite clearly by looking at them. Internally the 
actual values are measured much more often (every few 
milliseconds). They can always be read via an interface 
card, but they are still constantly changing and are not 
strictly assigned to level A or B and could thus also be 
measured during rise or fall.
6.7.4  Rise/fall time
The rise and fall times are equal and are thus conside-
red as one value. Also see figure 10. It is continuously 
adjustable in a range of 30µs...200ms. The step width is 
changing from µs to ms over the whole range. The error 
lies at max. +10%.
The time is displayed as slew rate in the format ∆value/
time. Example: the set value of A has been set to 40A 
and the one of B to 20A, then a current leap of 20A is 
generated in a time x, for instance 100ms. The display 
then shows 20A/100ms. The ∆value is not adjustable at 
this point.
Note:  the  pulse  widths  of A  and  B  should  always  be 
greater  than  the  rise/fall  time,  else  the  pulsed  signal 
would result in triangle-shaped wave form or something 
else. For example, a pulse width of 100ms for A and B 
and a rise/fall time of 100ms creates a triangle-shaped 
signal with a period time of 200ms.This might be wanted 
in certain applications and is thus not limited or influenced 
by the device.
Handling the device 
6.8  The battery test mode
Introduction
Mode „Battery“ is intended for battery tests, by connecting 
a battery to the load and discharge it definedly. The ave-
rage current is measured and the elapsed time is counted 
and then displayed as the consumed battery capacity in 
Ah. The voltage supervision, together with the adjustable 
undervoltage  shutdown  threshold 
Ulow
,  prevents  the 
battery  from  being  deeply  discharged. This  threshold 
needs to be adjusted at least once. If it is exceeded du-
ring the test, the load input is automatically switched off 
and the time counter is halted. No more current is drawn 
from the battery. If the threshold is set to greater than the 
battery voltage, the test can‘t be started.
Selecting the regulation mode
The  preselected  regulation  mode  determines  the  set 
values that can be adjusted for the test. The mode can 
be changed at any time,
even while the test is running. 
Doing so will reset the whole test, along with time counter 
and Ah value.
Usage
Before and during the test you can adjust the set value of 
the (pre)selected regulation mode (CC, CR or CP) and the 
undervoltage shutdown threshold 
Ulow
. The value to set 
is selected with 
Selection(5) and adjusted with Setting(6)
The displays shows the elapsed test time in the format  
Hours:Minutes:Seconds  (HH:MM:SS),  as  well  as  the 
consumed capacity in Ah.
Calculation of the Ah value
The ampere hours value (withdrawn electric charge) is 
calculated from the average of the last two measurings 
of the input current and the elapsed time.
Start/pause/stop the test
The test ist started by pressing the pushbutton 
Input on/
off(4) and either stops after 100 hours are counted or 
the battery voltage exceeds the undervoltage threshold. 
Another press of the button 
Input on/off(4) during the test 
halts it, another press will continue the test. A reset of the 
time counter is done by leaving the battery test mode by 
switching the selector 
Level Control(3) to position Level
A/B or Setup or by switching to a different regulation 
mode with 
Mode(2).
Note: if external control by analogue interface is activated 
(pin REMOTE = low) while the battery test is running, the 
test is aborted. After external control has left again and if 
the switch „
Level Control“ is still in position „Battery“, the 
test can be started again. Time and Ah value are reset.
Note: if the regulation mode is changed by Mode(2), all 
other set values that are not adjustable for the selected 
regulation mode are set to default values in order to let 
the test mode work correctly. Hence the setting 
Keep set 
values 
is ineffective here.
The example in figure 13 shows that the current of 100A 
can  not  be  reached  because  the  power  limitation  has 
become active (example of a 2400W model).
Note: The time display is not 100% exact. The deviation 
of the displayed time to the actually elapsed time can be 
1-2 seconds per hour.