Kern Counting scales Weight range 6 kg Readability 0.1 g mains-powered, rechargeable Silver CPB 6K0.1N Manual Do Utilizador
Códigos do produto
CPB 6K0.1N
CPB-N / CPB-DM-BA-e-1323
15
6.7 Adjustment
As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that there are no objects on the weighing plate.
Observe stable environmental conditions. A warming up time (see chapter 1) is
required for stabilization. Ensure that there are no objects on the weighing plate.
Provide calibration weight, for details, see chapter 1 "Technical Data. "
6.7.1
Models CPB-N (non verified models)
Operation
Display
Switch on balance and during self-check
press the TARE key
Use the numeric keys to enter password:
Default password "0000"
Should this password entry not be possible,
enter a personal password; (entry using
enter a personal password; (entry using
function
see. chap.11.1).
Also possible to continue the adjustment
process with
button.