Operating Manual V 1.1 – 05/2007
6.5.2 Installing the Storage Systems (Optional)
6.5.3  Shelf, Movable (Optional)
Attach the rod for the shelf to the inner front side
of the cryostat housing using the provided screws
(1) and the No. 3 Allen key, then attach the caps
(3). (The rear side of the movable shelf has white
plastic screws (2) that prevent the interior of the
chamber from being scratched.) Now hook the
movable shelf into the guide rod.
6.5.4 Inserting the Section Waste Tray
Fig. 22
6.
Installation
To do so, remove the insert (1), place the frame
(2) in front of the bore and tighten the screws/
washers on the cryostat housing using the No. 4
Allen key. Afterwards, insert the insert (1) into the
frame and fold it up.
Fig. 23
Before installing the knife/
blade holder base, insert the
section waste tray with the
cutout (a) facing the user.
a
1
2
Fig. 21
For reasons of
accessibility, the
(optional) storage
system must always
be installed first.
1
2
3