Teledyne Camera Accessories T100 Manual Do Utilizador
Calibration Procedures
Teledyne API - T100 UV Fluorescence SO2 Analyzer
190
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
ENTR
CONC
EXIT
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
CAL
SETUP
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
ZERO
CONC
EXIT
EXIT
returns to the main
SAMPLE
display
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
M-P CAL
SO2 STB=X.XXX PPB
SO2
=XXX.X
< TST TST >
SPAN
CONC
EXIT
M-P CAL RANGE = 500.0 PPB
SO2
=XXX.X
< TST TST >
ENTR
SPAN CONC
EXIT
Wait until SO2 STB
falls below 0.5 ppb.
This may take several
minutes.
The SPAN button now
appears during the
transition from zero to span.
You may see both the
SPAN and the ZERO
buttons.
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST TST >
CAL
SETUP
Set the Display to show the
SO2 STB
test function.
This function calculates the
stability of the SO
2
measurement
Press ENTR to changes the
OFFSET
& SLOPE values for the
SO
2
measurements.
Press EXIT to leave the calibration
unchanged and return to the
previous menu.
ACTION:
Allow span gas to enter the sample port at the
rear of the instrument.
The value of
SO2 STB
may jump
significantly.
Wait until it falls back
below 0.5 ppb.
This may take several
minutes.
Press ENTR to change the
OFFSET
& SLOPE values for the
SO
2
measurements.
Press EXIT to leave the calibration
unchanged and return to the
previous menu.
SAMPLE RANGE = 500.0 PPB SO2 =XXX.X
< TST
TST
>
CAL
SETUP
M-P CAL
RANGE = 500.0 PPB
SO2
=XXX.X
< TST TST > ENTR CONC
EXIT
Figure 9-3: Setup for Manual Calibration without Z/S valve or IZS Option (Step 3)
06807C DCN6650