Teledyne Camera Accessories T100 Manual Do Utilizador
Calibration Procedures
Teledyne API - T100 UV Fluorescence SO2 Analyzer
194
Step Three: Perform the calibration or calibration check according to the following
flow chart:
flow chart:
Wait until STABIL
falls below 0.05
ppb. This may
take several
minutes.
ZERO CAL M
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ZERO
CONC
EXIT
returns to the
main SAMPLE
display
Pressing ENTR changes the calibration of the instrument.
If either the ZERO or
SPAN
button fails to
appear, see Chapter 11
for troubleshooting tips.
Analyzer enters ZERO
CAL
mode.
EXIT
returns the unit to
SAMPLE
mode without
changing the calibration
values.
ZERO CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST > ZERO CONC
EXIT
ZERO CAL M STABIL=XXX.X PPB SO2=X.XX
X
< TST TST > CAL
CALZ
CALS
SETUP
SPAN CAL M
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
SPAN
CONC
EXIT
EXIT
returns to the
SAMPLE
mode without
changing the calibration
values.
ACTION:
Allow zero gas to enter the sample port at the
rear of the instrument.
SAMPLE RANGE = 500.000 PPB SO2 =XXX.X
< TST
TST >
CAL
CALZ
CALS
SETUP
SAMPLE
STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST > CAL
CALZ
CALS
SETUP
Scroll the display to the
STABIL
test function. This
function calculates the stability
of the SO
2
measurements.
The value of STABIL
may jump
significantly. Wait
until it falls below 0.5
ppb. This may take
several minutes.
Pressing ENTR changes the calibration of the instrument.
Analyzer enters SPAN
CAL
Mode.
SPAN CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ENTR
CONC
EXIT
SPAN CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
SPAN
CONC
EXIT
ZERO CAL M STABIL=XXX.X PPB
SO2 =XXX.X
< TST TST >
ENTR
CONC
EXIT
Figure 9-7: Setup for Manual Calibration with Z/S Valve Option Installed (Step 3)