Calibration Procedures 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
194 
Step Three: Perform the calibration or calibration check according to the following 
flow chart: 
 
Wait until STABIL 
falls below 0.05 
ppb. This may 
take several 
minutes. 
ZERO CAL M
       
 STABIL=XXX.X PPB
        SO2 =XXX.X  
 
 < TST  TST >  
ZERO
      CONC 
EXIT
 returns to the 
main SAMPLE 
display  
Pressing ENTR changes the  calibration of the instrument.
 
If either the ZERO or 
SPAN
 button fails to 
appear, see Chapter 11 
for troubleshooting tips. 
Analyzer enters  ZERO 
CAL 
mode
EXIT
 returns the unit to 
SAMPLE
 mode without 
changing the calibration 
values. 
ZERO CAL M       STABIL=XXX.X PPB  
       SO2 =XXX.X  
 
 < TST  TST >  ZERO      CONC
 
 
EXIT
 
ZERO CAL M        STABIL=XXX.X PPB         SO2=X.XX
 X 
 
 < TST  TST >  CAL
 
 CALZ   
CALS
 
 
SETUP
 
SPAN CAL M
       STABIL=XXX.X PPB
         SO2 =XXX.X  
 
 < TST  TST >  
SPAN
      CONC
 
 
EXIT
 
EXIT
 returns to the 
SAMPLE
 mode without 
changing the calibration 
values. 
ACTION:
Allow zero gas to enter the sample port at the 
rear of the instrument.
SAMPLE          RANGE = 500.000 PPB        SO2 =XXX.X   
 
< TST  
TST >
  CAL
 
 CALZ 
  
CALS
 
 
SETUP
 
 SAMPLE             
STABIL=XXX.X PPB
          SO2 =XXX.X  
 
< TST  TST >  CAL
 
 
CALZ
   
CALS
 
 
SETUP
 
Scroll the display to the 
STABIL
 test function. This 
function calculates the stability 
of the SO
2
 measurements. 
The value of STABIL
may jump 
significantly. Wait 
until it falls below 0.5 
ppb. This may take 
several minutes. 
Pressing ENTR changes the calibration of the instrument.
 
Analyzer enters  SPAN 
CAL
 Mode. 
SPAN CAL M        STABIL=XXX.X PPB
        SO2 =XXX.X  
 
 < TST  TST >  
ENTR
      CONC
 
 
EXIT
 
SPAN CAL M        STABIL=XXX.X PPB
         SO2 =XXX.X  
 
 < TST  TST >  
SPAN
      CONC
 
 
EXIT
 
ZERO CAL M        STABIL=XXX.X PPB   
      SO2 =XXX.X  
 
 < TST  TST >  
ENTR
      CONC 
 
EXIT 
 
Figure 9-7: Setup for Manual Calibration with Z/S Valve Option Installed (Step 3) 