Calibration Procedures 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
200 
 
 
Note 
The CALIBRATE attribute must always be set to OFF for analyzers used in 
US EPA controlled applications that have IZS option installed. 
Calibration of instruments used in US EPA related applications should 
only be performed using external sources of zero air and span gas with an 
accuracy traceable to EPA or NIST standards and supplied through the 
analyzer’s sample port (refer to Section 9.2). 
The following example sets Sequence 2 to carry out a zero-span calibration every other 
day starting at 01:00 on September 4, 2002, lasting 15 minutes. This sequence will start 
0.5 hours later each day. 
 
Table 9-4:  Example Auto-Cal Sequence 
MODE / ATTRIBUTE
VALUE 
COMMENT 
SEQUENCE 
Define Sequence #2 
MODE 
ZERO-SPAN 
Select Zero and Span Mode 
TIMER ENABLE 
ON 
Enable the timer 
STARTING DATE 
Sept. 4, 2002 
Start after Sept 4, 2002 
STARTING TIME 
01:00 
First Span starts at 01:00 
DELTA DAYS 
Do Sequence #2 every other day 
DELTA TIME 
00:30 
Do Sequence #2 0.5 h later each day 
DURATION 
15.0 
Operate Span valve for 15 min 
CALIBRATE 
ON 
The instrument will re-set the slope and offset 
values for the SO
2
 channel at the end of the AutoCal 
sequence 
06807C DCN6650