Principles of Operation 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
286 
13.1.7.2. PMT OPTICAL FILTER 
The PMT used in the T100 reacts to a wide spectrum of light which includes much of 
the visible spectrum and most of the UV spectrum. Even though the 214 nm light used 
to excite the SO
2
 is focused away from the PMT, some of it scatters in the direction of 
the PMT as it interacts with the sample gas. A second optical band pass filter placed 
between the sample chamber (refer to Figure 13-2) and the PMT strips away light 
outside of the fluorescence spectrum of decaying SO
2
* (refer to Figure 13-5) including 
reflected UV form the source lamp and other stray light.  
 