Teledyne Camera Accessories T100 Manual Do Utilizador

Página de 422
Principles of Operation 
Teledyne API - T100 UV Fluorescence SO2 Analyzer  
308 
13.5.7.3. STATUS LEDS & WATCH DOG CIRCUITRY 
Zero/Span and IZS O ptions  
    Sam ple/C AL Valve
 
IZS O ption 
Perm eation T ube H eater
 
D ark Shutter
 
Sam ple C ham ber H eater
 
I2C  W atchdog  LED
 
Zero/Span and IZS O ptions  
   Zero/Span Valve
 
 
Figure 13-21:  Relay Board Status LED Locations 
Thirteen LEDs are located on the analyzer’s relay board to indicate the status of the 
analyzer’s heating zones and valves as well as a general operating watchdog indicator. 
Table 13-1 shows the state of these LEDs and their respective functionality. 
Table 13-1: 
Relay Board Status LED’s 
LED 
COLOR 
FUNCTION 
STATUS WHEN LIT 
STATUS WHEN UNLIT 
D1 RED Watchdog 
circuit 
Cycles On/Off every 3 seconds under control of the CPU.
 
D2 YELLOW 
Sample chamber 
(RCELL) heater 
HEATING NOT 
HEATING 
D3, D4 
YELLOW 
Unused 
N/A 
N/A 
D5 
YELLOW 
IZS heater (option) 
HEATING 
NOT HEATING 
D6 
YELLOW 
Unused 
N/A 
N/A 
D7 
GREEN 
Zero / Span Valve 
Valve open to Span Gas path 
Valve open to Zero Gas 
(normal state) 
D8 
GREEN 
Sample / Cal Valve 
Valve open to  
calibration gas path 
Valve open to sample gas 
inlet on rear panel  
(normal state) 
D9, D10 
GREEN 
Unused 
N/A 
N/A 
D11 
GREEN 
UV Lamp Shutter 
Shutter open 
Shutter closed 
D12-14 
GREEN 
Unused 
N/A 
N/A 
06807C DCN6650