The Krown  100 can operate from battery power, making your unit completely portable. To use battery
power, unplug the AC charge/adapter from the unit.
When the batteries are fully charge, the Krown 100 will operate
for about 1/2 hours. The batteries automatically recharge whenev-
er the unit is plugged in to the AC charger/adapter and the unit is
turned OFF.
If you are using your Krown 100 on battery power and the red
Power Light starts to turn green, this shows that the battery needs
recharging. The light will be green for several minutes. Then the
Krown 100 will stop working. When this happens, you must
change over to AC power (house electricity) to continue using
your machine until you recharge the battery.
It is suggested that you exercise the battery in your Krown 100 at
least once a month. That is, operate the unit on battery power
until it needs recharging. Then recharge the battery by plugging in
the AC charger/adapter for 12 to 14 hours. (The unit must be
turned OFF in order to recharge.) Repeat this cycle at least once a
month. When used regularly in this cycle, the battery should last
for about three years.
The replaceable, sealed battery pack is a single unit that contains
the battery, connector and cover. This is located on the underside
of the Krown 100 at the back of the unit. Replacement battery
packs are available from Krown Manufacturing or your local
Krown distributor.
To replace the battery pack, depress the thumb notch in the center
of the cover and slide the unit to the back of the TDD. Carefully
unplug the connector. Replace with a new battery pack unit.
XIV
35