APPENDIX E:  Oscilloscope Use & Setup for Troubleshooting 
 
9.5.1 General 
Setup 
 
•  Connect Channel A to Pin 6 
•  Connect GND to Pin 3 
•  Connect Trigger to External to Pin 4 
•  Set Volts to 500 mV / div 
•  Set Time to 50 ųSec / div 
 
Your output should resemble this: 
Pin 5 
Not used 
Pin 1 
Not used 
Pin 2 
Not used 
Pin 3 
Ground or 
Common 
Pin 4 
Sync Pulse 
Pin 6 Return 
Signal 
9.5.2  Use and setup of the oscilloscope for the MT2000 
 
Prerequisites: 
•  Requires a Dual Channel Oscilloscope 
•  Procedure is based upon Fluke portable Oscilloscopes 
 
Scope Settings: 
 Channel 
A: 
•  Connect Probe A tip to “Signal 2” 
•  Set Range to 1 Volt / Div. 
•  Set input A to DC Coupled 
•  Connect probe grounds to “Common” 
 Channel 
B: 
•  Connect Probe tip to “Sync. Pulse”; Note:  “Sync Pulse” is actually the Blanking Signal and its position 
changes when BLK is changed. 
•  Set range to 1 volt/div 
•  Set input B to DC coupled 
•  Set time to ~500 uSec/Div. 
•  Change timing and trigger:  up and down as needed to see top fo probe or the whole probe length, de-
pending on overall length 
•  Set Trigger to Channel B, set level at 2 volts, “negative slope”, “wait for trigger”, “recurrent”, per scope 
options in trigger menu. 
•  Set time delay to –2 division 
 
NOTE:  The signal waveform will be visible on Channel A.  When changing the Blanking (BLK) the waveform will 
move.  Increasing BLK will shift the waveform to the left.  If the scope time delay is set to –2 Div., then set the blank-
ing so that any coupler pulses below 2 vdc are to the left of the end of the second division.  Set THV to a value that 
crosses the wanted signal, but does not cross any other baseline down going “bumps”.  
Figure 9.4