Command ‘c’ — Sub-command Index 05: Select Data in a Stream
 
 
This sub-command sets options that cause a specified stream to deliver specific kinds of 
information to host. By default, only Pressure EU data are delivered for the channels already 
specified by the “00” command. 
 
 
Command 
c 05 st bbbb” 
 
‘c’  is the command letter 
‘05’ is the sub-command index (‘ii’) for Select Data
st’ is the stream id digit (1, 2, or 3, (0 not allowed)  
‘ bbbb’ is the hex option field (bit map) to select which options will be 
returned in the data stream (see table) 
 
NOTE: all parameters are separated by a space character
 
Response 
A” 
‘A’ is the acknowledge letter 
 
Description:  If this sub-command is never executed for a particular stream, then Pressure EU  
 
 
Data are delivered (by default) in that stream following the fixed format binary  
  header 
( tssss as described by the “01” sub-command).  However, this sub- 
  command 
may 
also 
delete these default pressure readings from a stream (by not  
 
 
specifying them) as well as add other selected acquired data to a stream (by  
  specifying 
them). 
 
 
The bit map values (shown in the following table) may be added together to 
specify all the actual data groups that will be delivered in each packet of the 
specified stream.  The first two table entries, if their “bits” are specified,  will 
cause two-byte binary (16-bit, big endian) status values to be delivered in the 
stream packet (immediately following the binary stream header).  The third table 
entry, if specified, will cause the Pressure EU Data to be delivered (next), per 
the specified format (f), and for just the channels specified  in the configured 
stream. The remaining table entries  will cause other special data groups (i.e., 
raw pressures and EU temperature values, also in A/D counts or voltage forms) 
to also be delivered in each stream packet.  Each of these special data groups is 
also output, if its “bit” is specified, in the order of its table entry (within the 
packet).  Each group will also have a datum per the specified channels, and be in 
the specified format (per f). 
 
 
 
Note
 
Selecting too many other data groups will compromise module 
performance.