SECTION 8—ACCESSORIES
36
NOTE: Stable being defined as the weight fluctuating two tenths of a pound. For example, a 
patient weighing one hundred pounds, the scale will fluctuate between 99.8 and 100.2 until the 
LOCK key is pressed. Fluctuation of the weight displayed is normal as noted above. Press the 
LOCK button to lock the weight.
NOTE: The UNITS button can be pressed to toggle between units of pounds and kilograms. This 
is indicated by lb or kg appearing in the display window.
8. The lift may now be lowered and the sling removed from the patient.
NOTE: The patientʹs weight will continue to be seen in the display window. The display will turn 
off automatically after a two minute period of non‐use [no changes in weight exceeding five pounds 
(two kilograms)]. You can NOT adjust the time delay for automatic shut off. After the display has 
turned off, the weight may be recalled by pressing the ON/OFF button. The unit can be turned off 
by pressing the ON/OFF button a second time. 
Replacing the Battery
NOTE: For this procedure, refer to FIGURE 8.4.
NOTE: The scale is powered by a nine volt alkaline battery that should provide approximately 1500 
readings before needing replacement.
When battery replacement is needed, LO BAT will appear on the display. Perform the 
following:
1. Slide the battery door open in the direction of the arrow.
2. Remove existing battery.
3. Install the new battery.
4. Reinstall the battery door.
Calibrating the Reliant Scale
NOTE: The Reliant Scale will be pre‐calibrated at the factory with the load cell. Should it be 
necessary to re‐calibrate the scale, follow the instructions outlined below.
1. The patient and the sling must be removed from the scale to properly calibrate the 
Reliant Scale. For removing the patient instructions refer to Operation on page 14. For 
sling detachment instructions refer to Lifting the Patient on page 19.
2. With the Reliant Scale on, remove the four screws on the back of the enclosure and 
remove the front cover to expose the PC board.
Battery 
Door
9V Battery
Assembled 
Scale
Arrow