IDEAL INDUSTRIES Thermometer 61-633 Manual Do Utilizador

Página de 23
Page 13 
 
Testing the Voltage Function 
To verify accuracy in the DC and AC voltage ranges, do the following: 
1.   Connect the Calibrator to the V
Ω and COM inputs on the Meter. 
2.   Set the Calibrator for the voltage from step 1 to 5 in Table 1. 
      Compare the reading on the Meter display with the display reading shown in Table 1. 
      If the display reading falls outside of the range shown in Table 1, the Meter does not  
      meet specification. 
 
Table 1. DC Voltage Test 
Input Reading 
Step Voltage 
61-633 
61-635 
3.6000V 
3.5908 to 3.6092 
3.5958 to 3.6042 
-3.6000V 
-3.5908 to -3.6092 
-3.5958 to -3.6042 
36.000V 
35.908 to 36.092 
35.954 to 36.042 
360.00V 
359.08 to 360.92 
358.58 to 360.42 
900V 
896.2 to 903.8 
897.46 to 902.54 
 
3.   Press the blue button on the Meter to toggle to ACV function. 
4.   Set the calibration for the voltage and frequency from step 1 to 17 in Table 2. 
      Compare the reading on the Meter display with the display reading shown in Table 2. 
      If the display reading falls outside of the range shown in Table 2, the meter does not  
      meet specification. 
 
Table 2. AC Voltage Test 
Input 
Reading 
Step Voltage 
Frequency 
61-633 
61-635 
3.6000V 
50Hz 
3.5590 to 3.6410 
3.5698 to 3.6302 
3.6000V 
1KHz 
3.5230 to 3.6770 
3.5590 to 3.6410 
3.6000V 
10KHz 
3.4860 to 3.7140 
3.5220 to 3.6780 
3.6000V 
20KHz 
N.A 
3.4850 to 3.7150 
3.6000V 
50KHz 
N.A 
3.4120 to 3.7880 
36.000V 
50Hz 
35.518 to 36.482 
35.698 to 36.302 
36.000V 
1KHz 
35.230 to 36.770 
35.590 to 36.410 
36.000V 
10KHz 
34.320 to 37.680 
35.220 to 36.780 
36.000V 
20KHz 
N.A 
34.850 to 37.150 
10 
36.000V 
50KHz 
N.A 
34.120 to 37.880 
11 
360.00V 
50Hz 
355.18 to 364.82 
356.98 to 363.02 
12 
360.00V 
1KHz 
352.30 to 367.70 
355.90 to 364.10 
13 
360.00V 
10KHz 
348.60 to 371.40  
352.20 to 367.80 
14 
360.00V 
20KHz 
N.A 
348.50 to 371.50 
15 
360.00V 
50KHz 
N.A 
341.20 to 378.80 
16 
720.0V 
50Hz 
706.4 to 733.6 
709.96 to 730.04 
17 
720.0V 
1KHz 
699.6 to 740.4 
703.20 to 736.80 
 
 
Form Number TM61633-5 
Rev 5 Dec 2006