Table 12. Frequency Test 
Input Reading 
Step Frequency 
Level 
61-633 
61-635 
360Hz 
0.5Vp-p 
359.86 to 360.14 
359.86 to 360.14 
3.6000KHz 
0.5Vp-p 
3.5986 to 3.6014 
3.5986 to 3.6014 
36.000KHz 
0.5Vp-p 
35.986 to 36.014 
35.986 to 36.014 
360.00KHz 
0.5Vp-p 
359.86 to 360.14 
359.86 to 360.14 
3.6000MHz 
1.5Vp-p 
3.5986 to 3.6014 
3.5986 to 3.6014 
 
Testing the Duty Factor Function 
To verify the accuracy of the Meter’s Duty Factor function, do the following: 
1.   Connect the Calibrator to the V
Ω and COM inputs on the Meter. 
      Note   The accuracy of the Calibrator’s Duty function must be appropriate for the  
                 specified accuracy of the Meter. 
2.   Set the rotary switch to Hz. 
3.   Press the blue button on the meter to toggle to %DF function. 
4.   Set the Function Generator for square wave voltage for steps 1-3 of Table 13. 
     (Frequency is about 1KHz) 
5.   Compare the reading on the Meter display with the display reading shown in Table 13. 
6.   If the display reading falls outside of the range shown in Table 13, the Meter does not 
      meet specification. 
 
 Table 13. Duty Factor Test 
Input Readings 
Step Duty 
(%)  Level 
61-633 
61-635 
20.0 
5Vp-p 
14.0 to 26.0 
14.0 to 26.0 
50.0 
5Vp-p 
44.0 to 56.0 
44.0 to 56.0 
77.0 
5Vp-p 
71.0 to 83.0 
71.0 to 83.0 
 
Testing the Temperature Function 
To verify the accuracy of the Meter’s Temperature Function do the following: 
1   Connect the calibrator to V
Ω and via K-type wire and T-V adaptor 
     (Ideal 61-635 TC adapter) 
     Note: The T-V adaptor should be allowed to stabilize to the same room temperature as the 
      meter before beginning test. 
2.   Set the rotary switch to 
°C. 
3   Set the Calibrator for inputs in Table 14. 
4.   Compare the reading on the Meter display with the display reading shown in Table 14. 
5.   If the display reading falls outside of the range shown in Table 14, the Meter does not 
      meet specification. 
 
Table 14 Temperature Test 
Input Readings 
Step Temperature 
61-633 
61-635 
-50.0
°C 
N.A 
-52.1 to -47.9. 
-20.0
°C 
N.A 
-21.1 to -18.9 
0.0 
°C 
N.A 
-1.1 to 1.1 
20.0
°C 
N.A 
18.9 to 21.1 
50.0
°C 
N.A 
48.9 to 51.1 
100.0
°C 
N.A 
98.9 to 101.1 
500.0
°C 
N.A 
498.9 to 501.1 
1000
°C 
N.A 
998.9 to 1001.1 
 
Form Number TM61633-5 
Rev 5 Dec 2006