Hitachi Microscope & Magnifier S-4800 Manual Do Utilizador

Página de 294
 
4-
21
4.10  Gun Baking operation  
 
 CAUTION 
Gun baking operation must be carried out by a service engineer or a person who has been 
trained by Hitachi High-Technologies Corporation or its service representatives.     
 
Use the following steps for gun baking operation. 
 
(1)   Turn HV to OFF.   
 
NOTICE 
Confirm that the HV Lamp on the right side ion pump cover is 
extinguished. If it still is lit, never go to next steps. 
 
 
 
 
(2) Remove two fixing screws and then the protection cover.   
(A protection is provided so that gun HV is forced to off when the 
cover is removed. ) 
 
 
 
 
(3) Pull out the HV cable from the electron gun 
 
 
 
 
 
(4) Insert the HV cable into the HV cable holder provided at the left 
side of ion pump cover.   
(A protection is provided so that baking will not start if the cable 
is not placed in the holder.)   
 
 
 
 
(5) Remove two screws at the top of the column cover and remove 
the cover.   
 
 
 
HV Lamp
Protection cover 
Screws