1-10 
1.3.1    Power and Ground Wirings 
 
The instrument operates only with a power supply of AC100 V.     
If using a power voltage other than AC100 V, the double insulation transformer (option) is 
required.  
The transformer shall be put after MAIN switch of S-4800. Fig 1-3 shows connection of the step 
down transformer. 
 
 
 
 
 
 
 
 
 
 
 
Fig 1-3    Connection of the step down transformer 
 
To connect the transformer, remove the left side part of the rear cover of display unit.   
Remove the provided two short circuit cables put on the TB1 terminal of the power distribution 
unit.  
Connect the cable of input of the transformer to I(L) and 2(N) , and the cable of output of the 
transformer to 3(L) and 4(N) terminals of the TB1. Refer the following fig 1-4 for connection. 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
Fig 1-4    Connection of the step down transformer 
The cable of the step down transformer shall be prepared locally with the transformer. 
200V: 32A   
100V 
UL/CE:Double Insulation 
Safety circuit 
MAIN-ELB 
100/200V
Current 
Braker
Remove short cable 
Remove short cables 
To step down 
transformer  
From 
Transformer 
Fixing band for cable to transformer   
To step down 
transformer 
1(L) 2(N) 3(L) 4(N)
TB1 TRANS (OPTION)
Main Power
To Column 
Unit