Hitachi Microscope & Magnifier S-3400N Manual Do Utilizador

Página de 466
3.5.7 
3 - 75 
3.5.7.8  Position Memory Function 
 
The stage coordinates, both absolute and relative, can be memorized and retrieved. 
200 absolute coordinates (20 points on each of 10 pages) and 20 relative coordinates can be 
saved in total.    Three axes, X, Y, and R, can be registered. 
Click the Memory button on the Stage tab.    The following Position Memory dialog window will 
open.  
 
 
 
Fig. 3.5-49    Stage Memory Dialog Box 
 
(1)  Registration of absolute stage coordinates at the present stage position 
Select a Page and Number, then click Get Position.    The present coordinates are then 
indicated in the X, Y and R boxes.    Input any comments in the Comment box, and click 
Save. 
 
(2)  Registration of absolute stage coordinates by direct input 
Select a Page and Number, and input values to the X, Y and R boxes using the keyboard 
followed by the Enter key.    Input any comments in the Comment box, and click Save. 
 
(3)  Registration of relative stage coordinates using the present stage positions 
 
(a) Select page Rel and Number. 
 
(b)  Move the stage (using RISM) to the first position and click Get Relative-1. 
 
(c)  Move the stage to the second position and click Get Relative-2. 
The relative coordinates between the first and the second positions are calculated and 
indicated in X, Y and R boxes. 
 
(d)  Input comments in the Comment box and click Save.