Sono-Trak™ Doppler Ultrasonic Flow Meter
First Time Operation
Once you have successfully completed installation of the Doppler flow meter, you are
ready to configure your unit for operation using the front panel in the electronics
enclosure.
ready to configure your unit for operation using the front panel in the electronics
enclosure.
Table 2-3. LED conditions and indications
LED
Condition
Indication
Power
(Green)
Off
Power is not available to the unit.
If you have tried to power up your
unit and this light is off, see
unit and this light is off, see
.
Illuminated steady
Power is available to unit.
Proceed with normal operation.
Echo
(Yellow)
Off
No echo signal.
Consult
Flashing
Echo signal is weak.
Consult
Illuminated steady
Echo signal is good.
Proceed with normal operation.
Alarm
High or Low
(Red)
Off
No alarm conditions exist.
Proceed with normal operation.
Illuminated steady
An alarm condition exists.
Your system is running at high or low
flow. Adjust your flow levels or
consult
flow. Adjust your flow levels or
consult