Fujimoto Fuji Differential Pressure Transmitter FKK 4 Manual Do Utilizador

Página de 74
63
Pressure range of FKP
Smart type
Pressure range
[kPa] {bar} <psi>
8.125~130 {0.08125~1.3} <1.188~19>
31.25~500 {0.3125~5} <4.375~70>
187.5~3000 {1.875~30} <26.875~430>
625~10000 {6.25~100} <93.75~1500>
CALIBRATION
The transmitter should be calibrated in a calibration room.  For calibration of each transmitter, the
following devices are required.
o Pressure source and pressure measuring equipment (should have as high an accuracy as pos-
sible)
*  Measurable ranges are listed in the table below.
o  Power supply : DC power supply (24 V DC) or Fuji Electric FC series power supply unit
(type PXJ or PXL)
o Load resistor : Standard resistor 250 
Ω (within ±0.0125 Ω)
o Measuring device : Digital voltmeter (capable of measuring transmitter output with an
accuracy better than 0.1%)
*  Use meter having a 5-digit display.
o Hand Held Communicator (HHC) type FXW
*  For use with smart type
A2
Differential pressure range of FKK
Preparation for calibration
Smart type
Differential pressure range
[kPa] {mmH
2
O} <inH
2
O>
0.375~6 {37.5~600} <1.5~24>
2~32 {200~3200} <8~125>
8.125~130 {812.5~13000} <32.5~520>
31.25~500 {3125~50000} <4.375~70psi>
125~2000 {12500~200000} <17.5~280psi>
Pressure range of FKH
Smart type
Pressure range
[kPa] {bar} <psi>
8.125~130 {0.08125~1.3} <1.188~19>
31.25~500 {0.3125~5} <4.375~70>
187.5~3000 {1.875~30} <26.875~430>