Fluke Cable Analyzer DTX-ELT Manual Do Utilizador
Power
Main unit and remote: Lithium-ion battery pack, 7.4 V, 4000 mAh
• Typical battery life: 12 to 14 hours
• Charge time (with tester off): 4 hours (below 40 °C)
• AC adapter/charger, USA version: Linear power supply; Input 108 V ac to 132 V ac, 60 Hz ;
Output 15 V dc, 1.2 A
• AC adapter/charger, international version: Switching power supply; Input 90 V ac to 264 V ac,
48 to 62 Hz; Output 15 V dc, 1.2 A (isolated output)
• Memory backup power in main unit: Lithium battery
• Typical life of lithium battery: 5 years
• The battery will not charge at temperatures outside of 0°C to 45°C (32°F to 113°F).
The battery charges at a reduced rate between 40°C and 45°C (104°F and 113°F).
• Typical battery life: 12 to 14 hours
• Charge time (with tester off): 4 hours (below 40 °C)
• AC adapter/charger, USA version: Linear power supply; Input 108 V ac to 132 V ac, 60 Hz ;
Output 15 V dc, 1.2 A
• AC adapter/charger, international version: Switching power supply; Input 90 V ac to 264 V ac,
48 to 62 Hz; Output 15 V dc, 1.2 A (isolated output)
• Memory backup power in main unit: Lithium battery
• Typical life of lithium battery: 5 years
• The battery will not charge at temperatures outside of 0°C to 45°C (32°F to 113°F).
The battery charges at a reduced rate between 40°C and 45°C (104°F and 113°F).
Languages Supported
English, French, German, Spanish, Russian and Italian
Calibration
Service center calibration period is 1 year
Performance Specifications
Note: all specifications for tests on twisted-pair cabling apply to 100 Ω cable. Contact Fluke Networks
for information on measurement performance for cable with a different impedance.
for information on measurement performance for cable with a different impedance.
Baseline accuracy
The baseline accuracy of all DTX Series testers meets or exceeds the requirements of Level IV in
2nd Edition of IEC 61935-1 over the frequency range supported by the tester.
2nd Edition of IEC 61935-1 over the frequency range supported by the tester.
Category 6/Class E link
test modes (and lower link
categories)
DTX CableAnalyzer far exceeds Level III requirements of TIA/EIA-568-B.2-2 and IEC 61935-1
Length for Twisted Pair Cabling
(Length specs do not include the uncertainty of the cable’s NVP Value)
Parameter
(Length specs do not include the uncertainty of the cable’s NVP Value)
Parameter
Single-ended Test
Dual-ended Test (Main and Remote)
Range
800 m (2600 ft)
150 m (490 ft)
Resolution
0.1 m or 1 ft
0.1 m or 1 ft
Accuracy
± (1m + 4%)
± (1m + 4%)
DC Loop Resistance Test
Parameter
Parameter
Twisted Pair Cabling
Range
O Ω to 53Ω
Resolution
0.1 Ω
Accuracy
± (1 Ω + 1%)
Overload Recovery Time
Less than 10 minutes to rated accuracy following over-voltage. Referencing is required after repeated or
prolonged over-voltage.
prolonged over-voltage.
Delay Skew
Parameter
Parameter
Twisted Pair Cabling
Range
0 ns to 100 ns
Resolution
1 ns
Accuracy
± 10 ns