Adam Equipment Welding Consumables HCB J302 Manual Do Utilizador

Página de 44
 
 
 
 
 
 
 
 
33
 
© Adam Equipment Company 2013
 
 
External Calibration 
1)
 
Press  and  hold  the  [Air/Liquid]  and  [Density]  keys  at  the  same  time  and 
power on the balance. 
2)
 
Release both keys when the display shows unload.  
3)
 
Remove any weight from the pan assembly and press the [Tare] key when the 
stable sign is shown.  
4)
 
The  display  will  show  an  acceptable  mass  that  can  be  used  to  calibrate  the 
unit. You can change this value by pressing the [Air/Liquid] key to match the 
calibration mass you may have. 
5)
 
 The calibration masses that can be used are as follows:  
Model # 
HCB J123 
HCB J302 
HCB J1002 
HCB J2002 
Weight 1 
50g 
100g 
500g 
500g 
Weight 2 
70g 
200g 
700g 
1000g 
Weight 3 
100g 
300g 
1000g 
2000g 
 
6)
 
Once the calibration weight has been selected press the [Tare] key 
7)
 
The  display  shows  LoAd.  Place  your  weight  on  the  pan.  Press  the  [Tare]  key 
after the stable sign is shown.  
8)
 
The display will show PASS if the calibration is successful.  
9)
 
Remove the weight and the display will show zero. 
 
NOTE:  If  the  calibration  fails  try  doing  the  calibration  again.    The  calibration  must 
agree with the factory calibration within ±5%.   
The balance will show FAllllL H (when the weight is higher) or FAllllL L (when the weight 
is lower).  Repeat the process using the correct calibration weight.