GDC-150 Operation Manual 
 
 
 
5209-9000 Rev 0 
3.  INSTALLATION 
 
3.1.  Mounting Holes and Conduit Ports 
 
The  GDC-150  should be installed on a flat vertical surface with the sensor pointing 
outwards in a clean, dry environment.  If the GDC-150 is to be installed in a potentially 
wet environment, the optional water-tight enclosure should have been selected.  This 
reference refers to the  standard, general purpose PVC  enclosure.  Four mounting 
holes are provided in the enclosure base for securing the GDC-150  to the wall.  Do 
not block the front of the enclosure as this is where the sensor is situated and where it 
monitors the air for the target gas. 
 
 
NOTE:  Four mounting holes are provided in the enclosure base for 
securing the  GDC-150  to the wall.  The mounting holes are 3/16
″ (4.8 
mm) in diameter. 
 
Conduit entry points are provided for the PVC enclosure.   
 
 
Figure 3-1.  Standard Enclosure Showing Mounting Holes 
The water-tight enclosure must be installed with the mounting screws passing through 
the same openings that accommodate the four door securing screws.  This ensures 
the mounting screws are outside of the door gasket and confirms a water-tight 
installation. 
 
 
NOTE:  No conduit entry points are provided for the water-tight 
enclosure.  The reason is the installer may not want one where we 
would place it and this would result in unnecessary openings into this 
enclosure.