Braun IRT 1020 Manual Do Utilizador

Página de 10
 
Thank you for 
purchasing the 
Braun ThermoScan 
Instant Thermometer
 
Unlike other thermometers that must be held 
in place for several minutes, the  Braun 
ThermoScan Instant Thermometer takes a 
temperature in just one second. It takes a 
“picture“ of the infrared heat given off by the 
ear drum and surrounding tissue, calculates, 
and displays an ear temperature on the digital 
readout.
But way take a temperature at the ear? 
Because the eardrum is close to and shares 
blood vessels with the hypothalamus, the 
temperature control centre of the brain. 
Therefore, temperatures taken from the ear 
reflect the body’s “core“, or internal 
temperature. (Tis is the temperature of the 
heart, brain and internal blood supply.) The 
ear is also a protected cavity, so, unlike oral 
temperatures, the accuracy of ear 
temperatures in unaffected by eating, 
drinking, smoking and breathing through the 
mouth. The ear is a dry cavity so cross-
contamination through mucous membrane 
contact, a problem with oral and rectal sites, 
cannot occur. Also, the risk of bowel 
perforation, a concern when taking rectal 
temperatures on infants or children, is 
eliminated.
The Braun ThermoScan Instant Thermometer 
has been clinically proven to be safe and 
accurate. To ensure the best results, please 
take the time to read this manual completely 
and keep it handy for future reference.
4
5
 
IRT 1020 product specifications
 
Product type:
6005
Technical Characteristics:
Patient temperature range
34 - 42.2°C (93,2-108°F)
Surrounding room temperature 
for operational range
116 - 40°C (61-104°F)
Display resolution
0.1°C or °F
Temperature scales (user selectable)
°C or °F
Operating relative humidity (max)
80 %
Long term storage ranges
Temperature
–20 to 50 °C (– 4 to 122°F)
Humidity (max)
95 % noncondensing
Display Mode:
Ear ( )
Power source:
One 9 volt alkaline battery
(6 LR 61, MN 1604 9 V, alkaline)
Weight (without batteries):
188 g (6.6 oz)
Accuracy characteristics:*
Applicable patient ages
All ages
Laboratory error (max)
Patient temperature range
Error °C
Error °F
 
37.0°C to 39.0°C
(8.0°F to 102.0°F)
±0.1
±0.2
35.8°C to <37.0°C
(96.4°F to <98.0°F)
±0.2
±0.3
>39.0°C to 41.0°C
(>102.0°F to 106.0°F)
<35.8°C or >41.0°C (<96.4°F or >106.0°F) ±0.3
±0.5
 
* Meets American Society for Testing and Materials (ASTM) Accuracy 