Welch Allyn Medical Diagnostic Equipment Medical Alarms 8.1x Manual Do Utilizador

Página de 210
Directions for use
Chapter 10   Troubleshooting and maintenance
159
Printing
Table 18.  Troubleshooting printing
Problem
Explanation
There is a delay in the Virtual Monitor or 
Waveform Window data display when a 
printout takes place. Is any data lost?
No data is lost.
In snapshots or alarm prints, sometimes four 
or five waveforms are printed, and sometimes 
three waveforms are printed.
Although a Virtual Monitor can show up to three waveforms at once, if 
additional waveforms are turned on for display, all are printed. 
An alarm print or snapshot doesn't include a 
vital signs summary, even though an NIBP 
measurement was taken within the last eight 
hours.
When over three waveforms are printed on an alarm print or snapshot, 
there is no room left on the printout for a vital signs summary. 
If you prefer to receive a vital signs summary, turn off additional 
waveforms in the Virtual Monitor by clicking MAIN MENUMORE
WAVE SELNEXT (if necessary), CHANGE.
I expected a printout to occur at the central 
station, but it did not occur.
Or
 appears on the map at the central 
station area.
Check for various causes
1.  Ensure that the printer is turned on and supplied with paper and 
toner. 
2.  Ensure that the printer’s On Line light is illuminated.
3.  Click 
 or 
 on the map at the central station location to
open the Central Station Printers window. 
4.  If any print job has 
 beside it, the job is corrupted due to a
system configuration problem. 
(a) Click the job.
(b) Click Delete, and then click Delete Selection.
(c) Contact Welch Allyn Technical Services, if necessary.
5.  If 
 appears in the lower left corner of the window, 
click 
 to open the Printers View. 
6.  Check for a printer with a problem message, and fix the problem. 
Printers with a stop-sign icon have been disabled.
7.  If the printer was disabled, click the printer name, and then click 
Enable.
An alarm is suspended and an alarm print 
occurs. Then another vital sign alarm occurs 
for the patient, but no an alarm print occurs. 
If an alarm is suspended and the Acuity System generates an alarm 
print, and then a new alarm occurs for the patient during the 
suspension period, a new alarm print does not occur.