Chapter 5   Respond to alarms and alerts
87
Respond to specific alarms and alerts
This section describes a few of the Acuity System alarms and alerts and how to respond 
to them.
Respond to a DROPOUT alert
When there is a loss of bi-directional communication between a wireless monitor and the 
Acuity System, an audible DROPOUT equipment alert occurs at the monitor and at the 
Acuity Central Station. 
The monitor continues to provide local alarms and alerts, but alarms generated by the 
Acuity System, such as arrhythmia or ST analysis alarms, are not indicated, and you 
cannot access any Acuity System windows related to the patient. 
To respond to a DROPOUT alert
Take one of these steps:
Click SUSPEND.
The audible tone is suspended for 90 seconds,.
Click RESET.
Alert indications stop and the Virtual Monitor is in the blue, “patient intentionally 
disconnected” state. 
If the monitor resumes communication with the Acuity System, the Virtual Monitor, 
Waveform Window and patient icon return to the green, “patient monitoring OK” 
state.