Welch Allyn Medical Diagnostic Equipment Thermometer 690 Manual Do Utilizador
71221-0000 Rev. H
25
Appendix A Specifications
Ambient Operating Temperature Range
50.0
F to 104.0 F (10.0 C to 40.0C)
Operating Altitude
-557 ft. to +16,000 ft / -170m to +4877m
Patient Temperature Range
80.0
F to 110.0 F (26.7 C to 43.3 C)
Temperature Predict Time*
Oral:
Approximately 4–6 secs.
Adult axillary (18 years and older):
Approximately 12–15 secs.
Pediatric axillary (17 years and younger):
Approximately 10–13 secs.
Rectal:
Approximately 10–13 secs.
Approximately 4–6 secs.
Adult axillary (18 years and older):
Approximately 12–15 secs.
Pediatric axillary (17 years and younger):
Approximately 10–13 secs.
Rectal:
Approximately 10–13 secs.
Transport/Storage Temperature
-13°F to +131°F (-25°C to +55°C)
Calibration Accuracy
+/- 0.2
F (0.1 C) (Monitor Mode)
Humidity
15% to 95% non-condensing
Dimensions
8.46” x 3.18” x 2.43”
(215mm x 81mm x 62mm)
(215mm x 81mm x 62mm)
Weight
12.6 ounces/357 grams
Power
(3) 1.5Vdc AA batteries
Equipment Classification
Class 2 Device, Continuous Operation
Not AP or APG equipment
Not AP or APG equipment
IPXØ
Not protected against the ingress of water.
* Predict time depends on probe placement and patient condition. Predict times do not include
tissue contact-detection time or time needed to place the probe into the measurement site.