SureTemp® Plus Service Manual
1
Electromagnetic Immunity and Emissions Declaration
Immunity Standard
Emissions Summary
The SureTemp
®
Plus model 690 and 692 thermometers are intended for use in the
electromagnetic environment specified below. The customer or the user of the model 690 and
692 should assure that it is used in such an environment.
692 should assure that it is used in such an environment.
IEC Basic Standard
IEC 60601-1-2 2001
Electrostatic discharge IEC 61000-4-2
± 6kV
c
± 8kV
c
RF EMF Immunity IEC 61000-4-3
3/10 V/m
80 MHz-2.5GHz
80 MHz-2.5GHz
Magnetic field IEC 61000-4-8
3A/m
150kHz-80MHz
150kHz-80MHz
Guidance and Manufacturer’s declaration - electromagnetic emissions
Emissions test
Compliance
Electromagnetic
environment-guidance
environment-guidance
RF emissions
CISPR 11
CISPR 11
Group 1
Class B
Class B
The SureTemp
®
Plus model
690 and 692 Thermometers
are suitable for use in all
establishments, including
establishments that are
connected to the public
low-voltage power supply
network.
are suitable for use in all
establishments, including
establishments that are
connected to the public
low-voltage power supply
network.
Harmonic emissions
IEC 61000-3-2
IEC 61000-3-2
Complies
Voltage fluctuations
and flicker emissions
IEC 61000-3-3
and flicker emissions
IEC 61000-3-3
Complies