3-2
When the shutter is closed the generator will switch off, unless another shutter is still  open  in one of the
other tube shield positions. Alternatively. the generator can be  switched off manually by operating the
GENERATOR 0 push-button. All lamps will extinguish. Before the generator is switched off the power
to the X-ray tube should be reduced, by first turning the  mA control to minimum, and then the  kV  control.
Any generation of HT will be inhibited if the shutter does not fully close after a close shutter command.
Fig. 3.1.
Opening  the  window  shutter.
3.3.
SHUTTER CONTROL
For camera work it is usual  to control the exposure time of the film to X-rays by using a PW 1722 timer. Two
timers can be fitted: the  left-hand  timer will control the shutter of either window 1 or window 2, and the
other timer the shutter of window 3 or 4. This paragraph describes the automatic and manual control of the
shutter operation.
When the camera is installed on  its  bracket in front of the required exit window of the tube shield, open the
shutter disc and push the camera forward until the collimator enters the hole in the disc. Ensure that the
microswitch button on the tube shield is actuated by the pin or arm of the  camera.
NOTE:
The space between the front of the collimator and the exit opening of the tube shield should be kept
as  small  as possible to ensure  that radiation does  not escape. Take care  that the filter disc is free
to   move,  and is not damaged.