Thermo Scientific Revco Cryogenic Storage Freezers
Installation and Operation
3
6 Operation 
6.1 Control Panel Features
Figure 2.  Revco Ultima II Series Control Panel with 
Optional Backup System
Before the initial start up, take some time to become familiar 
with the controls on your freezer. Figure 2 illustrates the Revco 
Ultima II Cryogenic Freezer control panel.
Note:
Figure 2 shows all possible functions. These functions 
vary depending on the options. You may not have all of 
the functions on your equipment. The CO
2
 or LN
2
 
Backup System lights (20, 21, 22) represent status 
indicators for a built-in CO
2
 or LN
2
 backup system, 
which is optional.
1. Three position keyed Power On Switch.
2. Power On Indicator. This indicator lights when power is 
connected to the freezer.
3. Temperature Failure Indicator. This indicator lights when the 
freezer temperature deviates either above or below the alarm 
temperature settings.
4. Power Failure Indicator. This indicator lights when there is a 
power failure to the freezer (refer to Section 7 on page 7).
5. Life-Guard Alert Indicator. This indicator lights when the 
first stage compressor has gone into a cautionary state of 
operation to help prevent failure.
6. Check Fuse Light. This indicator lights when the surge 
protector on the main electronic panel is blown by a power 
surge (refer to Section 7 on page 7).
7. Alarm Test Pad. Press this pad to start an alarm test (refer to 
Section 5.2.5 on page 6).
8. Alarm Reset Pad. Press this pad to reset slowly blinking 
indicators. Slowly blinking indicators denote that the 
condition has occurred but is now within the given operating 
parameters. This pad is also depressed to temporarily silence 
the audible alarm (refer to Section 5.2.4 on page 5).
9. Cold Alarm Setpoint. Press this pad to set the cold alarm 
(refer to Section 6.2.3 on page 4).
10. Warm Alarm Setpoint. Press this pad to set the warm alarm 
(refer to Section 6.2.3 on page 4). 
11. Control Setpoint. Press this pad to set the cabinet temperature 
(refer to Section 6.2.2 on page 4).
12. Cabinet Temperature Indicator. This indicator lights when the 
temperature display window is showing the cabinet 
temperature (refer to Section 6.2.1 on page 4).
13. Digital Temperature Display Window. This window displays 
chamber temperature, alarm values, etc. depending on the 
operating status of the freezer and the procedure being 
performed.
14. Decrement Pad (
). Use this pad to decrease temperature 
values.
15. Increment Pad (
). Use this pad to increase temperature 
values.
16. Extreme Ambient Alert. This indicator lights when the 
ambient temperature has exceeded the upper limit of the 
recommended operating range. If the Extreme Ambient Alert 
is activated, the ambient environment needs to be improved 
for proper function of the freezer.
17. Clean Filter Indicator. This indicator lights when the air 
cooled condenser filter is dirty (refer to Section 7 on page 7).
18. Alarm Battery Low Indicator. This indicator lights when the 
charge on the alarm battery is low. Press this button to 
display the percent of full battery charge in the digital display 
window.
19. Voltage Boost Indicator. This indicator lights when incoming 
voltage is low and the Voltage Safeguard has been activated 
(refer to Section 7 on page 7). Press this pad to display the 
current line voltage in the digital display window.
23. Alarm On/Set Point Security Indicator. This indicator lights 
when the alarm is activated.Also indicates that Setpoint 
Security is activated.
The following indicators are available only when the freezer is 
equipped with the optional backup system (refer to Section 6.3.4 
on page 6).
20. Backup System Supply On Indicator.
21. Backup System Empty Indicator
22. Backup System Battery Low Indicator.
off
life-guard alert
extreme ambient alert
clean filter
power failure
temperature failure
alarm testalarm reset
cabinet
temperature
control
set point 
alarm
battery low
voltage 
warm alarm
set point
cold alarm
set point
power on
alarm on
set point security
Ultima
1
15
7
9
10
11
SUPPLY
ON
SUPPLY
EMPTY
BACKUP SYSTEM
BATTERY
LOW
8
23
6
12
13
21
20
22
14
18
19
17
16
2
5
4
3
check
 fuse
boost