GE ISOVOLT Titan E Brochura

Página de 2
GE Inspection Technologies
Xray
GEInspectionTechnologies.com
53.5 (2.1")
40˚
∅ 
110 (4.4")
∅ 
100 (4")
304 (12")
506 (20")
0
20
40
60
80
100
120
140
160
1
4
2
3
5
6
Tubenhousings
Wire IQI [%]
Material thickness Al [mm]
:  160 MM2 (
geometr. magnification 6:1)
:  160 M2 
(with image enhancement)
:  160 M2
40
35
30
25
20
15
10
5
0
40 60
0
20
80 100 120 140 160 180
mA
kV
We reserve the right to technical modifications without prior notice.
©2004 General Electric Company. All rights reserved.
   GEIT-30131GB (04/05)
The generation of radiation in an X-ray 
tube solely depends on the operation 
values, not on the make. 
The dose rate relevant in practice and 
suitable for calculations of radiation 
protection values is defined by national 
standards; thus the dose rate of the tube-
housing ISOVOLT 160 M2/0.4-0.4HP, 
measured at a distance of 1 m from the 
focal spot, amounts to 2.5 Sv/h at max-
imum tube voltage and maximum anode 
dissipation.
This value must not be used to assess 
biological effects.
The dose rate of the leakage radiation is  
< 2.5 mSv/h (250 mrem/h).
Dose Rate within the Central Beam
Technical Data
Maximum tube voltage 
160 kV
 
Focal spot 1 
Focal spot 2
Maximum anode dissipation 
1000 W 
1000 W
Tube current at max. tube voltage 
6 mA 
6 mA
Focal spot size (EN 12 543) 
1.00 mm (~ 0.4 IEC 336) 
1.00 mm (~ 0.4 IEC 336)
Emergent beam angle 
40°
Inherent filtration 
1 mm Be
High voltage connection 
Plug socket for rubber cone plug R24 with optional quick-lock cable flange
Cooling water flow rate 
min. 4 l/min
Cooling water temperature 
max. 35° C
Cooling water pressure 
max. 6 bar
Weight (with optional cable quick-lock) 
8.5 kg (18.7 lbs)
Dimensions 
see drawing