Chapter 5 
Analog Measurements - Screens and Control Fields
Sensitivity (RF analyzer)
Sensitivity (spectrum analyzer)
Sensitivity (RF analyzer)
RF analyzer sensitivity adds about 6 dB of sensitivity for the ANT IN 
port when 
High
 is selected and the 
Input Atten
 field is set to 0 dB. 
Operating Considerations
Selecting 
High
 sensitivity may cause spectrum analyzer measurements 
to be uncalibrated when the ANT IN port is used (a message appears 
when this occurs).
High-level AM measurements may be distorted when high sensitivity is 
used with the ANT IN port. 
Screens on which this field is displayed
RF ANALYZER
Sensitivity (spectrum analyzer)
This area performs two functions: 
• The lower field selects the vertical resolution of the display. You can 
choose from 1 dB, 2 dB, or 10 dB per graticule.
• The upper field selects 
Normal
 or 
High
 sensitivity for the RF input. 
The 
High
 setting adds about 6 dB of sensitivity to the ANT IN port 
(when the 
Input Atten
 field is set to 0 dB) for looking at very low 
level signals. However, this setting can cause measurements to be 
uncalibrated (indicated by a message on the screen). 
High
 
sensitivity can also cause high-level AM signals to be distorted. 
Screens on which this field is displayed
SPEC ANL: auxiliary controls menu